HP 16530A Service Manual page 65

Digitizing oscilloscope module
Table of Contents

Advertisement

HP 16530/31 A- Adjustments
w.
Touch Calibration, then touch Channel. Set s/Div to 5 ns.
x.
Touch Channel, then touch Calibration.
y.
Touch label of next channel to calibrate. See figure 4-17.
( Oscilloscope C) (Calibration}
( Autoscale
J
~
(
Cal Choice
Ch_ch_skew
(
A2o
s) ~- (
Figure 4-17. Channel for Skew Calibration
Stop
)
B2 0
s)
z.
Repeat steps u through y until all channels are calibrated to the reference channel.
4-12
.-.......

Advertisement