HP 16530A Service Manual page 46

Digitizing oscilloscope module
Table of Contents

Advertisement

HP 16530/31A- Performance Tests
g.
Set Auto markers screen up as shown in figure 3-16.
Oscilloscope D
J (
Displey
( Autoscele)
~
(
)( ->
(
Set on
l
(
et Level
C1
50
1
II
(
Set on
J
(
et Level
)(
C1
50
1
0
(
S te tis tics
)
On
(
Run Until Time
)(-Q
J
)(
Off
0
Done
Figure 3-16. Auto Markers Configuration
Marl{er X: Set on channel 1 at 50% level - Positive edge - on 1
Marker 0: Set on channel 1 at 50% level - Positive edge - on 2
Statistics: On
h.
Touch Done. Touch Run, then drag finger to Repetitive.
i.
Read the X to 0 value. The limits are 397.2 ns to 403.8 ns.
j.
Change frequency on signal generator to 100 MHz.
lc
Change timebase to 5 ns/Div.
I.
Reacl the X to 0 value. The limits are 9.15 ns to 10.85 ns.
Run
3-11