HP 16530/31A- Performance Tests
f. Configure Trigger screen as shown in figure 3-4.
Osc 1 II a scope D ) (
Tr~
gger
)
(
~~~:
l (
So~~ce
] (
Lev~
I
v]
Semple Period
40.000 ns
Figure 3-4. Trigger Screen Configuration
Mode: Edge
Source: Channel 1
Level: 0 V
Count:
1
Auto-Trig: Off
g. Configure Display screen as shown in figure 3-5.
Semple Period
40.000 ns
Figure 3-5. Display Screen Configuration
Mode: Average
Avg-no: 4
Connect the dots: Off
h.
Touch Display. then touch Auto-Meas.