HP 16530/31 A- Adjustments
e. Touch channel input label of next channel to be calibrated and change impedance to 50
Ohms. See figure 4-7.
Figure 4-7. Channel to be Calibrated
f.
If there are additional acquisition cards, repeat steps d and e until all channels are set to 50
Ohms.
g.
Touch input label field as shown in figure 4-8.
Figure 4-8. Input Label Field
h.
Touch Channel Mode, then touch Overlay.
4-7
-------------------------------~~-----