HP 16530f31A- Performance Tests
i.
Touch Autoscale , then touch Continue, as shown in figure 3-6 and record the value of
Vp-p.
Figure 3-6. Auto-Meas the Vp-p
j.
Change frequency of signal generator to 100 MHz.
k.
Touch Autoscale , then touch Continue and read the value of Vp-p. This value should be
~
70% of the value recorded in step i.
I.
Change amplitude on signal generator to 600 mV and change the frequency to 200 kHz.
m.
Repeat steps i through k.
n.
Set signal generator to 200 kHz with an amplitude of 300 mv.
o.
Repeat steps i through n. testing all other channels, and specifying each channel as the
trigger and measurement source.
- - - - - - - - - - - - - - - - - - - - -
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
- - - - - - - -
--
3-5