HP 16530A Service Manual page 63

Digitizing oscilloscope module
Table of Contents

Advertisement

HP 16530/31 A - Adjustments
p.
Using two cables of equal length and a BNC tee, connect Compensation Signal as shown
in figure 4-13 to 1 N PUT 1 and IN PUT 2 of the top acquisition card. IN PUT 1 of the top ac-
quisition card will be the reference for all other channels. INPUT 2 will be the first channel
to be calibrated.
HP16531A ACQUISITION
HP16531A ACQUISITION
0-
INPUT 1
HP16531A ACQUISITION
0
-INPUT 1
INPUT 2
,...,..,0
TEE
Figure 4-13. Channel-To-Channel Skew Equipment Hookup
q.
Touch Autoscale, then touch Continue.
r.
Touch Run, then drag finger to Repetitive. See figure 4-14.
Cel Chotce
Ch_ch_ske"'
Figure 4-14. Run Option Select
s.
Touch Calibration, then touch Channel. Set s/Div to 5 ns.
t.
Touch Channel, then touch Calibration.
4-10