Hayward HydroRite UVO3 Troubleshooting Manual page 25

Hide thumbs Also See for HydroRite UVO3:
Table of Contents

Advertisement

Cleaning Quartz Sleeve & Lamp Sensor (cont.)
DO NOT remove the lamp sensor from the vessel body. Pro Tip: When reattaching
the stainless steel sleeve, first align the quartz sleeves with the concave returns
(looking through the clear plastic bottom piece), then flip the whole unit over to
verify the sleeves are lined up with the bezels before securing.
Step 13
Step 14
Next, clean the lamp sensor with isopropyl
Once the cleaning procedure is complete AND
alcohol (rubbing alcohol). The lamp sensor
the quartz sleeves are completely dry, reverse
should be cleaned in this fashion every time
steps 1-10 to reassemble the unit. NOTE: the
the quartz sleeves are maintenance. Once
stainless steel sleeve only goes on in one
cleaned proceed to step 14.
orientation (DO NOT force it).
*NOTE: After cleaning is complete DO NOT Calibrate the HydroRite, this should
ONLY be done when all three lamps are replaced.