10.3 Fuse

10.4 Test Points

Centra-W Cell Washer Service Manual
There is one fuses on the circuit board, labeled F1, to protect the logic
circuitry. A blown F1 fuse will result in no display. Test the fuse with a meter
to determine if replacement is necessary. The fuse is rated as follows:
F1
6.3 A, 125 V 5x20mm Slo-Blo IEC Part No. 8976
The board contains 3 test points. They are as follows:
TP1
GND
TP2
+5V
TP3
10 Hz
Ground
Logic Voltage
Frequency