Interacoustics AT235 Service Manual page 6

Impedance audiometer
Table of Contents

Advertisement

Service Manual
AT235
The panel can now be carefully removed by lifting
it up in the top end and releasing the bottom from
its holder.
Fig. 3
Releasing the module from the instrument allows
easier access. This is done by carefully releasing
the tube on the right hand side.
Fig. 5
The silicone tube is easily removed from the
pump outlet by pulling it apart gently. Pushing the
pump arm slightly, the spring is ready to be
unhooked.
Fig. 7
The pump module which is attached to the access
panel can now be flipped over. Please note that it is
still attached with a cable and a tube.
Fig. 4
By sliding the black lock to the right allowing release
of the cable, the module is no longer attached to the
instrument.
Fig. 6
It is now possible to open the pump arm to the left,
and the module is ready to be de-attached.
Fig. 8
4