Service Manual
AT235
The panel can now be carefully removed by lifting
it up in the top end and releasing the bottom from
its holder.
Fig. 3
Releasing the module from the instrument allows
easier access. This is done by carefully releasing
the tube on the right hand side.
Fig. 5
The silicone tube is easily removed from the
pump outlet by pulling it apart gently. Pushing the
pump arm slightly, the spring is ready to be
unhooked.
Fig. 7
The pump module which is attached to the access
panel can now be flipped over. Please note that it is
still attached with a cable and a tube.
Fig. 4
By sliding the black lock to the right allowing release
of the cable, the module is no longer attached to the
instrument.
Fig. 6
It is now possible to open the pump arm to the left,
and the module is ready to be de-attached.
Fig. 8
4