Interacoustics AT235 Service Manual page 30

Impedance audiometer
Table of Contents

Advertisement

Service Manual
AT235
CIR 33
4
Equipment needed for calibrating the CIR 33
Measuring amplifier
Acoustical IEC 60318-5 (IEC 126) coupler, fitted with eartip
10
Frequency counter
Correction values for coupler
From the AUD menu, select Transducer Calibration.
Place the transducer on the 2cc IEC 60318-5 (IEC 126) coupler
(fig. 46)
DD65 V2
5
Equipment required:
Measuring amplifier
Acoustical IEC 60318-3 (IEC 303) coupler
Frequency counter
Correction values for coupler
Spacer for HDA200 to coupler IEC 60318-3 (IEC 303)
From the AUD menu, select Transducer Calibration.
Place the transducer on the 6cc IEC 60318-3 (IEC 303) coupler
(fig. 47)
4 As of 2020, CIR33 is discontinued as a AD629 accessory. Hence, this information is applicable for previously purchased transducers only
5 Please note that this set-up does not refer to DD45 or DD65 v1
Fig. 46
Fig. 47
28