9.
Ensure that FAT32 is selected as your file system - Leave other settings as listed.
10. Click start - depending on the size of your thumb drive this may take a moment. When the formatting
is complete you will receive a pop-up indicating successful formatting.
11. Copy the update_user.logo.bin file to the formatted thumb drive.
12. It is very important that this file is the ONLY FILE on this USB thumb drive.
13. With the audiometer switched off, insert the thumb drive into any available USB port.
14. Then switch the instrument on and push the Temp/Setup Button from the Tone test screen.
15. Enter Common Settings using the Setup/Tests Button.
16. To the question "Do you want to install?" select Yes.
17. After the installation is completed, select Back to get to the testing screen.