Maintenance and Disassembly

2.1

General cleaning and maintenance procedure

CAUTION
Before cleaning the device, always make sure that the device is switched off and disconnected from
the power supply
Use a soft cloth lightly dampened with cleaning solution to clean all exposed surfaces
Do not allow liquid to come into contact with the metal parts inside the earphones / headphones
Do not autoclave, sterilize or immerse the instrument or accessory in any fluid
Do not use hard or pointed objects to clean any part of the instrument or accessory
Do not let parts that have been in contact with fluids dry before cleaning
Rubber ear-tips or foam ear-tips are single use components
Recommended cleaning and disinfection solutions:
Warm water with mild, nonabrasive cleaning solution (soap)
Procedure
Clean the instrument by wiping outer case with a lint free cloth lightly dampened in cleaning solution
Clean cushions and patient hand switch and other parts with a lint free cloth lightly dampened in
cleaning solution
Make sure not to get any moisture in the speaker portion of the earphones and similar parts