Reference Values For Stimulus Calibration - Interacoustics AT235 Service Manual

Impedance audiometer
Table of Contents

Advertisement

Service Manual
AT235
6.2

Reference values for stimulus calibration

Freq.
Reference Equivalent Threshold Sound Level
(RETSPL)
[dB re. 20 µPa]
[Hz]
TDH39
125
45
250
25.5
500
11.5
1000
7
1500
6.5
2000
9
3000
10
4000
9.5
6000
15.5
8000
13
WB
-8
LP
-6
HP
-10
*All figures in bold are Interacoustics Standard values.
EARtone
DD65 v2
3A / IP30
26
30,5
14
17
5.5
8
0
4,5
2
2,5
3
2,5
3.5
2
5.5
9,5
2
21
0
21
-5
-8
-7
-6
-8
-10
DD45 Probe NB
Stimulus
Correction
Values
47.5
41
4
27
24.5
4
13
9.5
4
6
6.5
6
8
5
6
8
12
6
8
11
6
9
3.5
5
20.5
3
5
12
-5
5
-8
-5
-6
-7
-10
-8
Variation of Ipsi
Sound
stimulus levels
attenuation
for different
values for
volumes of the
TDH39/DD45
ear canal
earphones using
Relative to the
MX41/AR or
calibration
PN51 cushion
performed on an
[dB]
IEC 126 coupler
[dB]
0.5 ml
1 ml
3
5
9.7
5.3
7
9.7
5.3
15
21 (1600 Hz)
11.7
3.9
26
-0.8
-0.5 31 (3150 Hz)
-1.6
-0.8 32
26 (6300 Hz)
24
7.5
3.2
8.0
3.6
3.9
1.4
V

Advertisement

Table of Contents
loading

Table of Contents