There are seven keys for menu navigation. The MENU key is used to toggle between the menu
and the measure screen. Pressing the Menu key provides options for calibration, configuration,
communications, outputs and relays. Refer to the AV38 Manual for Details.

3. INSTALLATION

3.1 System Mounting and Installation

The AquaTrace measurement system is mounted on a 12 inch X 12 inch panel with four ¼" mounting
holes. Use ¼ inch bolts with washers to mount the panel to a wall.
The sample supply is connected to the fitting at the bottom of the flow chamber. The sample supply
should be regulated to control pressure and flow. The sample pressure should not exceed 60 psi. The
sample flow rate should be regulated between 50 mL/min and 400 mL/min. The sensor will operate
properly across this range, however rapid variations in flow can cause temporary measurement
fluctuations. To insure a stable measurement, the sample supply should be regulated to provide a
consistent flow rate to the sample chamber.
The outlet fitting on the sample chamber must be connected to a vented drain. Excessive tubing length
or pipe restrictions that result in back pressure to the outlet are not allowed.
The inlet and outlet port on the sample chamber are supplied with compression fittings for use with ¼
inch OD tubing. When the inlet and outlet are to be directly plumbed, remove the compression fittings
and connect to the flow chamber with 1/8 inch NPT fittings
Note that the tubing and plumbing used to install the AquaTrace system must not be gas measuring the
process. Verify that all plumbing permeable. Stainless steel, Kynar PVDF and polypropylene are
relatively impermeable to oxygen. Other materials such as silicone are gas permeable and allow oxygen
contamination that will prevent the system from accurately components are suitable for use with a low
oxygen process.
Thermo Scientific AquaSensors
DataStick sensors connected to the AV38 communicate via MODBUS
RTU. As such, the AV38 can be used to select any two of the DataStick
sensors on the bus for display, current output reporting and relay alarms by
selecting the desired station addresses. When there is no DataStick
sensor connected at the selected network address, the measure screen
will indicate "DataStick Absent".
AquaTrace
Trace DO System User Guide
10