Thermo Scientific AquaSensors AquaTrace User Manual page 20

Trace do system
Table of Contents

Advertisement

8. The AV38 will display the calibration value for editing. Use the arrow keys to edit the
9. The AV38 will calibrate the temperature sensor to match the value entered. During this time, the
10. After the calibration point has been stored, the AV38 will prompt the user to confirm that the
11. When the enter key is pressed, the analog output is placed into active mode and the monitoring
12. If an error occurs during calibration that causes the procedure to fail, the reason for the failure
Thermo Scientific AquaSensors
temperature value to the known value of the calibration bath. When the value is correct, press
the ENTER key to continue.
MENU
calibration procedure can be aborted by pressing the ESC key.
MENU
calibration procedure has been successful. Press the ENTER key to accept and continue.
MENU
of sensor and temperature values is resumed. This successfully completes the temperature
calibration procedure.
will be shown.
AquaTrace
Temperature ?
(
25.0)
ESC
ENTER
Temperature
Calibrating...
ESC
ENTER
Temperature
Confirm Cal OK
ESC
ENTER
Trace DO System User Guide
18