Figure 3-21. Using The Tc± Inputs On J5 As General-Purpose Signal - Texas Instruments ADS1x48EVM User Manual

Evaluation modul
Table of Contents

Advertisement

www.ti.com
3.5.4 Connecting a General-Purpose Input to J5 on the ADS1x48EVM
Although the ADS1x48EVM is designed for measuring all types of analog temperature sensors, general-purpose
input signals can be applied to the ADS1x48 through the TC± inputs on the J5 terminal block.
the input structure used for the TC± inputs.
Figure 3-21. Using the TC± Inputs on J5 as General-Purpose Signal Inputs
As
Figure 3-21
shows, resistors R24 and R23 are not populated by default, resulting in a set of analog inputs
that are only subject to the differential filter cutoff frequency of 1.93 kHz and the common-mode filter cutoff
frequency of 38.63 kHz. The TC± inputs can therefore be used for any general input signal as long as the
ADS1x48 absolute input and common-mode range requirements are satisfied. Select the AIN4 and AIN5 inputs
on the ADS1x48 to measure a voltage connected to the TC± inputs.
SBAU378A – SEPTEMBER 2021 – REVISED JANUARY 2022
Submit Document Feedback
AVDD
R24
DNP
10.0M
TC+
TC+
1
TC-
2
REF5025
3
4
REFOUT
NTC+
5
NTC-
6
TC-
J5
R23
DNP
10.0M
GND
Copyright © 2022 Texas Instruments Incorporated
R13
4.12k
C3
1000pF
C10
GND
0.01uF
C4
R16
1000pF
4.12k
ADS1x48EVM Overview
Figure 3-21
shows
AIN4
AIN5
ADS1x48EVM Evaluation Module
27

Advertisement

Table of Contents
loading

Table of Contents