Section 3: Adjustment
Command parameters for a dc current 10 A range adjustment
When calibrating your DMM6500 for a dc current 10 A range adjustment, use the following command
parameters. Connect a 10 A cable for this adjustment.
Send each command parameter twice. First, send the parameter using the setup command. Second,
send the parameter using the execute command.
The example code below uses the first line of code from the table below. Send all command
parameters using this command.
cal.adjust.step.setup("cal_DCI_10A_fs_pos")
cal.adjust.step.execute("cal_DCI_10A_fs_pos")
Command parameters
cal_DCI_10A_fs_pos
cal_DCI_DIGI_10A_fs_pos
cal_DCI_10A_fs_neg
cal_DCI_DIGI_10A_fs_neg
AC voltage adjustment
The following procedure provides instructions for completing a front-terminal adjustment for ac
voltage. The following section provides a command parameter table to complete the adjustment.
For this adjustment, you need a:
•
Fluke 5720 calibrator
•
Keithley Model 3390 function generator
3-26
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Calibrator
Calibrator
function
range
dc current
2.2
dc current
2.2
dc current
2.2
dc current
2.2
Calibrator
value
2
2
-2
-2
DMM6500-905-01 Rev. E June 2021