Adjustment Overview; Recommended Test Equipment - Tektronix KEITHLEY DMM6500 Calibration Manual

6 1/2-digit multimeter with scanning
Table of Contents

Advertisement

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual

Adjustment overview

DMM6500 adjustment is performed using a remote connection through either an optional GPIB, LAN,
or USB interface. The calibration adjustment commands provided in this manual use the Test Script
Processor (TSP
You can use the Keithley Test Script Builder to send you adjustment commands. See "Test Script
Builder (TSB)" in the Model DMM6500 Reference Manual (DMM6500-901-01).
To use GPIB with your DMM6500, you must use the KTTI-GPIB Communication and Digital I/O
Accessory.

Recommended test equipment

The following table summarizes the recommended calibration equipment. Specified accuracy of all
functions and ranges is dependent on the precision of reference signals used during the adjustment
process. To achieve specified performance, adjustment reference uncertainties must be at least four
times smaller than the best corresponding DMM6500 one-year accuracy specification for measuring
that signal.
Manufacturer
Fluke
Fluke
Fluke
Keithley
Instruments
Keithley
Instruments
Refer to the manufacturer's specifications to calculate the uncertainty, which varies for each function
and range test point.
DMM6500-905-01 Rev. E June 2021
) command language. There is no front-panel method for full adjustment.
®
Model
Description
5720A or 5730A
High Performance
Multifunction Calibrator
5725A
Amplifier
8508A or 8588A
8.5 Digit Reference
Multimeter
3390
Function/Arbitrary
Waveform Generator
8610 or 8620
4-Wire DMM Shorting
Plug
Section 3: Adjustment
Used for:
DCV, 10 A DCI, ACV,
ACI, and 10 kΩ
resistance
DCI and ACI 10 A
ranges
DCI
Frequency
DCV, digitize DCV, and
resistance
3-3