Digitize Current 10 A Range Verification - Tektronix KEITHLEY DMM6500 Calibration Manual

6 1/2-digit multimeter with scanning
Table of Contents

Advertisement

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 2: Performance verification

Digitize current 10 A range verification

Ensure cabling is designed to conduct 10 A without significant voltage drop. In general,
cables for the 10 A range steps should be made of heavy gauge wire and should be as short
as practical.
To verify the 10 A range:
1. Connect DMM6500, calibrator, and amplifier as shown in the following figure.
Figure 19: Connections for the 10 A range
DMM6500-905-01 Rev. E June 2021
2-49