DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Section 2: Performance verification
To prepare the DMM6500 for dc current accuracy verification:
1. Set up the DMM6500 for dc current and the range being tested. Make sure relative offset is
disabled.
2. Connect the calibrator, DMM6500, and reference DMM as shown in the following figure.
Figure 11: Connection for dc current
DMM6500-905-01 Rev. E June 2021
2-31