Tektronix KEITHLEY DMM6500 Calibration Manual page 26

6 1/2-digit multimeter with scanning
Table of Contents

Advertisement

Section 2: Performance verification
Verify the digitize voltage 100 mV range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify the digitize voltage 1 V range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify the digitize voltage 10 V range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify the digitize voltage 100 V range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify the digitize voltage 1000 V range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
2-16
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Nominal value
Lower limit
100
99.94
50
49.95
-50
-50.05
-100
-100.06
Nominal value
Lower limit
1
0.9996
0.5
0.49975
-0.5
-0.50025
-1
-1.0004
Nominal value
Lower limit
10
9.996
5
4.9975
-5
-5.0025
-10
-10.004
Nominal value
Lower limit
100
99.96
50
49.975
-50
-50.025
-100
-100.04
Nominal value
Lower limit
1000
999.6
500
499.75
–500
-500.25
–1000
-1000.4
Upper limit
100.06
50.05
-49.95
-99.94
Upper limit
1.0004
0.50025
-0.49975
-0.9996
Upper limit
10.004
5.0025
-4.9975
-9.996
Upper limit
100.04
50.025
-49.975
-99.96
Upper limit
1000.4
500.25
-499.75
-999.6
DMM6500-905-01 Rev. E June 2021