Tektronix KEITHLEY DMM6500 Calibration Manual page 87

6 1/2-digit multimeter with scanning
Hide thumbs Also See for KEITHLEY DMM6500:
Table of Contents

Advertisement

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Command parameters
cal_TS6
10 µA range (+) full scale nominal
cal_DCI_10uA_fs_pos
cal_DCI_DIGI_10uA_fs_pos
10 µA range (-) full scale nominal
cal_DCI_10uA_fs_neg
cal_DCI_DIGI_10uA_fs_neg
100 µA range (+) full scale nominal
cal_DCI_100uA_fs_pos
cal_DCI_DIGI_100uA_fs_po
s
100 µA range (-) full scale nominal
cal_DCI_100uA_fs_neg
cal_DCI_DIGI_100uA_fs_ne
g
*Characterized step measured by the Fluke 8508A or 8588A
Command parameters
1 mA range (+) full scale nominal
cal_DCI_1mA_fs_pos
cal_DCI_DIGI_1mA_fs_pos
1 mA range (-) full scale nominal
cal_DCI_1mA_fs_neg
cal_DCI_DIGI_1mA_fs_neg
10 mA range (+) full scale nominal
cal_DCI_10mA_fs_pos
cal_DCI_DIGI_10mA_fs_pos
DMM6500-905-01 Rev. E June 2021
Calibrator
Calibrator
Calibrator
function
range
value
n/a
n/a
n/a
dc current 0.00022
0.00001
dc current 0.00022
0.00001
dc current 0.00022
-0.00001
dc current 0.00022
-0.00001
dc current 0.00022
0.0001
dc current 0.00022
0.0001
dc current 0.00022
-0.0001
dc current 0.00022
-0.0001
Calibrator
Calibrator
Calibrator
function
range
value
dc current 0.0022
0.001
dc current 0.0022
0.001
dc current 0.0022
-0.001
dc current 0.0022
-0.001
dc current 0.022
0.01
dc current 0.022
0.01
Calibrator
Reference
stimulus value*
DMM dc
current range
n/a
8508A or 8588A
10 µA
Reading
8508A or 8588A
10 µA
Reading
8508A or 8588A
10 µA
Reading
8508A or 8588A
10 µA
Reading
8508A or 8588A
100 µA
Reading
8508A or 8588A
100 µA
Reading
8508A or 8588A
100 µA
Reading
8508A or 8588A
100 µA
Reading
Calibrator
Reference
stimulus value*
DMM dc
current range
8508A or 8588A
1 mA
Reading
8508A or 8588A
1 mA
Reading
8508A or 8588A
1 mA
Reading
8508A or 8588A
1 mA
Reading
8508A or 8588A
10 mA
Reading
8508A or 8588A
10 mA
Reading
Section 3: Adjustment
3-23

Advertisement

Table of Contents
loading

Table of Contents