Section 4: TSP command reference

cal.adjust.step.setup()

This function sets up the specified adjustment step.
Type
TSP-Link accessible
Function
Yes
Usage
cal.adjust.step.setup(stepname)
cal.adjust.step.setup(stepname, value)
stepname
value
Details
This command generates an error if:
Calibration is locked
The step does not complete successfully
The value that is passed is invalid for the step, out of range, or not needed
Also see
cal.lock()
(on page 4-6)
cal.unlock()
(on page 4-9)
4-4
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Affected by
The adjustment step to start
The value for this adjustment step. If using the default then value is optional
Where saved
DMM6500-905-01 Rev. E June 2021