Initial Instrument Setup; Select The Correct Terminals; Select The Tsp Command Set - Tektronix KEITHLEY DMM6500 Calibration Manual

6 1/2-digit multimeter with scanning
Hide thumbs Also See for KEITHLEY DMM6500:
Table of Contents

Advertisement

DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual

Initial instrument setup

Before adjusting calibration, make sure that the instrument is set up:
For remote operation and TSP commands.
To the correct date and time.
You must also unlock calibration.

Select the correct terminals

On the DMM6500, you must adjust calibration from both the front and rear terminals. You can verify
calibration on either the front or rear terminals. To set the instrument to the rear-panel terminals,
press the TERMINALS switch on the front panel of the instrument.

Select the TSP command set

Calibration adjustment must be performed by remote control using LAN, USB, or optional GPIB
interfaces. No front-panel calibration commands are available.
Calibration is only available using TSP commands. See the instructions below to change the
command set.
To set the command set from the front panel:
1. Press the MENU key.
2. Under System, select Settings.
3. For Command Set, select TSP.
4. You are prompted to reboot.
To verify which command set is selected from a remote interface:
Send the command:
*LANG?
To change to the TSP command set from a remote interface:
Send the command:
*LANG TSP
Reboot the instrument.
DMM6500-905-01 Rev. E June 2021
Section 3: Adjustment
3-5