Verify digitize current 100 µA range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify digitize current 1 mA range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify digitize current 10 mA range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
Verify digitize current 100 mA range
Description
Full scale (+)
Half scale (+)
Half scale (–)
Full scale (–)
2-36
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Calibrator setpoint
Lower limit
(A)
99.998
99.878
49.9991
49.9141
-49.993
-50.0843
-99.9985
-100.1185
Calibrator setpoint
Lower limit
(A)
0.999993
0.998993
0.499998
0.499348
-0.500009
-0.500659
-1.000017
-1.001017
Calibrator setpoint
Lower limit
(A)
10.00004
9.99204
5.00004
4.99454
-5.00003
-5.00553
-10.00006
-10.00806
Calibrator setpoint
Lower limit
(A)
100
99.92
50
49.945
-50
-50.055
-100
-100.08
Upper limit
100.118
50.0841
-49.9143
-99.8785
Upper limit
1.000993
0.500648
-0.499359
-0.999017
Upper limit
10.00804
5.00554
-4.99453
-9.99206
Upper limit
100.08
50.055
-49.945
-99.92