Ac Voltage Specifications - Tektronix DMM4040 User Manual

Table of Contents

Advertisement

4050 Accuracy
Accuracy is given as
(% measurement + % of range)
24 Hour
Range
(23 1 C)
100 mV
0.0025 + 0.003
1 V
0.0018 + 0.0006
10 V
0.0013 + 0.0004
100 V
0.0018 + 0.0006
1000 V
0.0018 + 0.0006
4040 Accuracy
Accuracy is given as
(% measurement + % of range)
24 Hour
Range
(23 1 C)
100 mV
0.003 + 0.003
1 V
0.002 + 0.0006
10 V
0.0015 + 0.0004
100 V
0.002 + 0.0006
1000 V
0.002 + 0.0006
Additional Errors
Digits

AC Voltage Specifications

AC Voltage specifications are for ac sinewave signals >5 % of range. For inputs from 1 % to 5 % of range and <50 kHz,
add an additional error of 0.1 % of range, and for 50 kHz to 100 kHz, add 0.13 % of range.
Maximum Input ....................................................... 1000 V rms or 1414 V peak or 8 x 10
Measurement Method ............................................. AC-coupled true-rms. Measures the ac component of input with up to
AC Filter Bandwidth:
Slow .................................................................... 3 Hz – 300 kHz
Medium ............................................................... 20 Hz – 300 kHz
Fast ..................................................................... 200 Hz – 300 kHz
Common Mode Rejection ....................................... 70 dB at 50 Hz or 60 Hz 0.1 % (1 k unbalance)
Crest Factor Error (applies to non-sinusoidal waveforms only)
Maximum Crest Factor ....................................... 5:1 at Full Scale
Additional Crest Factor Errors (<100 Hz)............ Crest factor 1-2, 0.05 % of full scale
90 Days
1 Year
(23 5 C)
(23 5 C)
0.0025 + 0.0035
0.0037 + 0.0035
0.0018 + 0.0007
0.0025 + 0.0007
0.0018 + 0.0005
0.0024 + 0.0005
0.0027 + 0.0006
0.0038 + 0.0006
0.0031 + 0.001
0.0041 + 0.001
90 Days
1 Year
(23 5 C)
(23 5 C)
0.004 + 0.0035
0.005 + 0.0035
0.003 + 0.0007
0.004 + 0.0007
0.002 + 0.0005
0.0035 + 0.0005
0.0035 + 0.0006
0.0045 + 0.0006
0.0035 + 0.0010
0.0045 + 0.0010
NPLC
Additional NPLC Noise Error
100
10
1
0.001 % of range
.2
0.0025 % of range +12 V
0.02
0.017 % of range +17 V
less) for any range.
1000 V dc bias on any range.
Crest factor 2-3, 0.2 % of full scale
Crest factor 3-4, 0.4 % of full scale
Crest factor 4-5, 0.5 % of full scale
Introduction and Specifications
Temperature
Coefficient/ C
Outside 18 to 28 C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
Temperature
Coefficient/ C
Outside 18 to 28 C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0 % of range
0 % of range
7
volts-Hertz product (whichever is
1
Electrical Specifications
1-17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Dmm4050

Table of Contents