Section 2: Performance verification
Verify dc voltage zero values using the 4-wire short
To verify dc voltage zero values:
1. Leave the short connected as described in
page 2-45).
2. Press the FUNCTION key.
3. Select the DC Voltage function.
4. Press the HOME key.
5. Set the range to 1000 V.
DC voltage verification is done in descending range order, starting with the 1000 V range and
finishing on the 100 mV range.
6. Verify that the 1000 V range zero is within specification. See the table below.
7. Verify that the 100 V to 100 mV range zero is within specification.
Verify dc voltage zero values
Range
1000
100
10
1
100 mV

Rear-panel verification

The following topics describe verification procedures that are done with connections attached to the
terminals on the DMM6500 rear panel.
2-46
DMM6500 6½-Digit Multimeter with Scanning Calibration and Adjustment Manual
Lower limit
Upper limit
-6000
6000
-600
600
-50
50
-6
6
-3.5
3.5
Verify resistance zero values using a 4-wire short
DMM6500-905-01 Rev. E June 2021