OMG100N1.21
5.9.2.6

Logging

Data logging mode can be started or stopped via 'Key 6 - Logging' accessed from the
'Utilities' menu. Press 'Key 4 – Start/Stop Logging' to start and stop data logging.
Whilst in data logging mode the instrument will automatically record data at the pre-set
intervals including running the pump for a preset time.
Active logging mode is indicated on the 'Main Read Screen' by the
The operator is able to edit the default ID, pump run-time, interval and start/stop
logging.
1)
From the 'Main Read Screen' press the 'Menu' key on the instrument panel.
2)
Press 'Key 6' to display the 'Logging' menu and the following screen is displayed:
3)
Select the desired option by pressing 'Keys 1 to 4'. Then enter the appropriate
setting using the keypad followed by the 'Enter' key.
1-Every 00 mins.
2-Pump 00 secs.
3-ID 00000000
4-Start/Stop Logging - Press 'Key 4' to start and stop data logging.
Note:
Page 22
Logging Menu
- Press 'Key 1' to enter the time in minutes for the
timeframe between sample readings. The interval
controls the reading frequency in minutes, i.e. every 10
minutes.
- Press 'Key 2' to enter the time in seconds for the length
of time you wish the pump to run when taking a sample
reading. The pump run-time is the time in seconds for
which the pump runs prior to the reading being stored.
This figure will also need to take into account the length
of sample tube and the volume of the sample gas. For
example, there is little point setting a pump run-time of
10 seconds if it takes 30 seconds to draw in a new
sample.
- Press 'Key 3' to create an 8-digit numeric ID Code.
Data logging mode is automatically stopped when the instrument is
switched off or if the logging parameters are edited.