G100 Range
4)
Use the 'Scroll' keys to adjust the value displayed. The larger the value, the less
sensitive the flow fail detection is.
5)
Press soft-key 'Accept' or 'Reject' accordingly.
Flow Fail Set-up Process – Best Practice
Copyright Geotechnical Instruments (UK) Limited
START
Decrease
the
No
displayed
value by 1
No
Wait 5 seconds
Flow fail
alarm?
Increase
the
displayed
value by 3
Turn pump
on
No
Block gas inlet
Wait 5 seconds
Flow fail
alarm?
Yes
Press pump
cancel
alarm
Decrease
the
displayed
value by 1
Accept
END
Flow Fail Set-up Process – Best Practice
G100 Range
Increase
the
displayed
value by 2
Press pump
Turn pump
cancel
on
alarm
Wait 5 seconds
Flow fail
alarm?
Yes
Decrease
the
displayed
value by 1
Block gas inlet
Wait 5 seconds
Flow fail
alarm?
No
Yes
Blocked or
excessive
restriction
Refer to Section 8.0
Problem Solving
OMG100N1.21
Page 21