Service

The G100 instrument should be regularly serviced to ensure correct operation and accurate
readings. The manufacturer recommends a full service and recalibration every 12 months.
Depending on usage the O
User Serviceable Parts
Note: There are no user serviceable parts inside the instrument. Please do not attempt
any repair as this may invalidate any warranty supplied with your instrument.
The following parts are supplied by your instrument manufacturer and can be user serviced:
Sample Filter
Sample Tubing
Cleaning
The instrument and accessories (including power supply unit) can be wiped clean using a
non-fibrous damp cloth.
Note: Do NOT apply pressure to the LCD display as this can cause damage.
Do NOT apply any moisture to the moisture removal tube on the rear of the
instrument as this may damage the membranes.
Do NOT use solvents or any other chemical cleaners.
Page 40
cell should be replaced every 2-3 years.
2
This should be regularly inspected for damage or discolouration
and changed if needed. The instrument should never be operated
without the sample filter as this may result in water or dust
entering the instrument. The filter should be changed immediately
if water can be seen. Failure to do so can damage the instrument.
Always ensure that sample tubes are not contaminated or
damaged.