6
Using the centrifuge
→
6.1
Initial start-up
DANGER
→
6.2
Switching the centrifuge on
→
6.2.1
Opening and closing the lid
WARNING
Version 05/2021, Rev. 1.2 of 28/04/2022 • sb
Translation of the original operating manual, part no. 0703502
•
Before the initial start-up, please ensure that your centrifuge is
properly set up and installed (see chapter 5 - "Set-up and
connection").
•
Press the mains power switch.
The display then illuminates. The centrifuge is ready for operation.
The lid can be opened if the centrifuge is at a standstill and if the lid key is
illuminated.
•
Press the lid key in order to open the lid.
The centrifuge cannot be started if the lid is opened.
•
To close, press with both hands slightly on the lid until both locks are
audibly locked.
Do not place your fingers between the lid and the housing when closing
the lid. Risk of crushing!
Blood bank centrifuge Sigma 8KBS
6 Using the centrifuge
33 / 125