4
Chapter 4 Remote Interface Reference

Pulse Configuration Commands

Pulse Configuration Commands

See also "Pulse Waveforms" starting on page 68 in chapter 3.
This section describes the low-level commands used to program the
function generator to output a pulse waveform. To select the pulse
function, use the FUNC PULS command (see page 171). Refer to the
diagram below for the command descriptions that follow.
90%
50%
Pulse Width
10%
Rise Time
Period
PULSe:PERiod {<seconds>|MINimum|MAXimum}
PULSe:PERiod? [MINimum|MAXimum]
Set the period for pulses. Select a period from 200 ns to 1000 seconds.
The default is 1 ms. MIN = 200 ns. MAX = 1000 seconds. The PULS:PER?
query returns the period of the pulse waveform in seconds.
• The specified period must be greater than the sum of the pulse width
and the edge time. The function generator will adjust the edge time
and the pulse width as needed to accommodate the specified period.
From the remote interface, a "Settings conflict" error will be generated.
The edge time is minimized first, and then the width (or duty cycle) is
adjusted as shown below.
Period > Pulse Width + (1.6 X Edge Time)
or in terms of pulse duty cycle:
Period > (Period X Duty Cycle ÷ 100) + (1.6 X Edge Time)
184
90%
50%
10%
Fall Time