Basic Measurements
Making Measurements

Setting Up Synchronized Pulse (IQ + Analog) Modulation

Ensure the equipment and triggers are properly configured. Refer to
"Equipment Setup" on page 9
Signal Analyzer" on page 11
You will need to create your own ASCII text marker file and a binary file with
the exact same name. Both files must be stored in the same file folder. For this
example, we will use example files that have been stored on the VXG.
Using the graphical user interface
On the VXG:
1. Select Preset > Preset to set the instrument to a known state.
2. Set the Frequency to 1 GHz and Power to 0.0 dBm.
3. Select the Signal block to open.
4. Select the arrow for Signal 1 to open the Signal Setup window.
This screen is only accessible if Option M9484C-8SG (8 virtual signal
generators) is installed. For all other option configurations, continue to the
next step.