Frequency Sweep Commands; Pulse Width Modulation (Pwm) Commands - Keysight 33210A User Manual

10 mhz function/arbitrary waveform generator
Hide thumbs Also See for 33210A:
Table of Contents

Advertisement

Frequency Sweep Commands

See also "Frequency Sweep" starting on page 90 in chapter 3.
Sweep Overview
The following is an overview of the steps required to generate a sweep.
The commands used for sweep are listed on page 205.
1 Select the waveform shape, amplitude, and offset.
Use the APPLy command or the equivalent FUNC, FREQ, VOLT,
and VOLT:OFFS commands to select the function, frequency, amplitude,
and offset. You can select a sine, square, or ramp waveform (pulse, noise,
dc, and arbitrary are not allowed).
2 Select the frequency boundaries of the sweep.
You can set the frequency boundaries using one of two methods.
a Start Frequency/Stop Frequency: Use the FREQ:STAR command to
set the start frequency and the FREQ:STOP command to set the
stop frequency.
To sweep up in frequency, set the start frequency < stop frequency.
To sweep down in frequency, set the start frequency > stop frequency.
b Center Frequency/Frequency Span: Use the FREQ:CENT command
to set the center frequency and the FREQ:SPAN command to set the
frequency span.
To sweep up in frequency, set a positive span.
To sweep down in frequency, set a negative span.
3 Select the sweep mode.
Use the SWE:SPAC command to select linear or logarithmic spacing for
the sweep.
Chapter 4 Remote Interface Reference

Frequency Sweep Commands

4
203