Chapter 5 Error Messages

Calibration Errors

Calibration Errors
The following errors indicate failures that may occur during a calibration
procedure (see chapter 4 in the Keysight 33210A Service Guide).
Calibration error; security defeated by hardware jumper
The function generator's calibration security feature has been disabled
by temporarily shorting the two "CAL ENABLE" pads on the internal
circuit board, as described in the Keysight 33210A Service Guide.
Calibration error; calibration memory is secured
A calibration cannot be performed when calibration memory is secured.
To unsecure the instrument, use the CAL:SEC:STAT ON command with
the correct security code.
Calibration error; secure code provided was invalid
Invalid security code specified with the CAL:SEC:STAT ON command.
Calibration error; provided value is out of range
The calibration value specified with the CAL:VAL command is out of range.
Calibration error; signal input is out of range
The internal analog-to-digital converter (ADC) has determined that the
signal applied to the rear-panel Modulation In connector is out of range.
707: Calibration error; cal edge time: rise time cal error
707: Calibration error; cal edge time: fall time cal error
707: Calibration error; cal edge time: default values loaded
Indicates a failure in the rise-time or fall-time circuitry has prevented
calibration. The edge time was calibrated using default values, limiting
accuracy. See the Keysight 33210A Service Guide for service information.
Calibration error; setup is invalid
You have specified an invalid calibration setup number with the
CAL:SET command. Refer to the Keysight 33210A Service Guide for
more information on the calibration procedures.
Calibration error; setup is out of order
Certain calibration setups must be performed in a specific sequence
in order to be valid. Refer to the Keysight 33210A Service Guide for
more information on the calibration procedures.