Chapter 2 Front-Panel Menu Operation

To Configure the Remote Interface

To Configure the Remote Interface

The Keysight 33210A supports remote interface communication using a
choice of three interfaces: GPIB, USB, and LAN (LXI Class C compliant).
All three interfaces are "live" at power up. The following sections tell how
to configure the remote interface from the instrument front panel.
Note: Two CDs, provided with your instrument, contain connectivity
software to enable communications over the remote interfaces. See
"Connectivity Software and Product CDs," page
information on these CDs and the software they contain.
GPIB Configuration
You need only select a GPIB address.
1 Select the "I/O" menu.
and then press the I/O softkey.
Press
2 Set the GPIB address.
Use the knob and cursor keys or the numeric keypad to select a GPIB
address in the range 0 through 30 (the factory default is "10").
The GPIB address is shown on the front-panel display at power-on.
3 Exit the menu.
Press the DONE softkey.
USB Configuration
The USB interface requires no front panel configuration parameters.
Just connect the Keysight 33210A to your PC with the appropriate USB
cable. The interface will self configure. Press the Show USB Id softkey in
the "I/O menu" to see the USB interface identification string. Both USB
1.1 and USB 2.0 are supported.