Communication Settings; Serial Interfaces; Protocol; Ascii-Protocol - B+B Sensors DM-Compact Series Operating Manual

Table of Contents

Advertisement

OPEARTING MANUAL
7.2

Communication Settings

7.2.1

Serial Interfaces

Baudrate
9,6...115,2 kBaud (adjustable on the unit or via Software)
DM 4M: 115,2 or 921,6 kBaud (adjustable on the unit or via
software)
Data bits
8
Parity
NONE
Stop bits
1
Flow controll
OFF
7.2.2

Protocol

All sensors of the DM series are using a binary protocol. Alternatively, they can be switched to an ASCII protocol (only LT versions). To get a fast
communication the protocol has no additional overhead with CR, LR or ACK bytes.
7.2.3

ASCII-Protocol

The models LT21, LT151, LT201, LT21H and LT101H can be switched to ASCII by changing the first figure of block 3 of the head calibration code.
This figure has to be changed from 0 to 4 (old sensing head) or 8 to C (new sensing head). [►4.3 Exchange of the Sensing Head]
Example (new Sensing head): Binary-Protocol:
Technische Änderungen vorbehalten
0141 0315-100
24.06.2022
A6FG – 22KB – 8AS0
1.Block 2.Block 3.Block
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
ASCII-Protokoll:
A6FG – 22KB – CAS0
1.Block 2.Block 3.Block
79 / 88

Advertisement

Table of Contents
loading

Related Products for B+B Sensors DM-Compact Series

This manual is also suitable for:

LtLtfLth1m2m3m ... Show all

Table of Contents