OPEARTING MANUAL
4.3
Exchange of the Sensing Head
From factory side the sensing head has already been connected to the
electronics and the calibration code has been entered. Inside a certain model
group any exchange of sensing heads and electronics is possible. The sensing
heads and electronics of the DM Fast models (LT151F and LT251F) and DM 4M
cannot be exchanged.
4.3.1
Entering of the Calibration Code
Every head has a specific calibration code, which is printed on the head cable. For a
correct temperature measurement and functionality of the sensor this calibration code
must be stored into the electronic box. The calibration code consists of 3 blocks (1M, 2M,
3M = 5 blocks) with 4 characters each.
Example :
A6FG
1.Block
For entering the code please press the Up and Down key (keep pressed) and then the
Mode key. The display shows HCODE and then the 4 signs of the first block. With Up and Down each sign can be changed; Mode switches to the
next sign or next block.
The entering of a new calibration code can also be made via the CompactConnect software (optional)..
Technische Änderungen vorbehalten
0141 0315-100
24.06.2022
–
22KB
–
0AS0
2.Block
3.Block
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
Bei Montage eines neuen Messkopfes
muss der Kalibriercode des neuen Kopfes
in die Elektronik eingegeben werden.
57 / 88
Need help?
Do you have a question about the DM-Compact Series and is the answer not in the manual?
Questions and answers