B+B Sensors HM 1150 Instruction Manual

Digital quick-response thermometer

Advertisement

Quick Links

INSTRUCTION MANUAL
Digital Quick-Response Thermometer
HM 1150
Description
Technical changes reserved
B+B 0141 0315-117 September 2010
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
01 / 12

Advertisement

Table of Contents
loading

Summary of Contents for B+B Sensors HM 1150

  • Page 1 INSTRUCTION MANUAL Digital Quick-Response Thermometer HM 1150 Description 01 / 12 B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen Technical changes reserved B+B 0141 0315-117 September 2010 Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com...
  • Page 2: Foreword

    Dear customer, We thank you for having purchased the Digital Quick-Response Thermometer HM 1150 and we are very glad that you deci- ded a product of B+B Thermo-Technik GmbH. We hope this product will fully satisfy you and will assist you effectively in your work.
  • Page 3: Table Of Contents

    INSTRUCTION MANUAL Content Foreword General Information Symbols Employed Warning Signs Safety Instructions Intendend Use Disposal 1. Points to be observed during operation 2. Recalibration 3. Correctiontable: (typ. value) 4. Specification 5. Scope of Delivery 6. Optional Order Information General Questions 03 / 12 B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen Technical changes reserved...
  • Page 4: Symbols Employed

    INSTRUCTION MANUAL Symbols Employed Notice Sign Meaning It is necessary to read the following advices before using the product. Advice The used symbols in the manual acts first of all as eye catcher for secu- rity risks. The symbols do not replace the security advices. The text must be read completely.
  • Page 5: Safety Instructions

    INSTRUCTION MANUAL Safety Instructions Für Schäden, die durch Nichtbeachten dieser Sicherheitshinweise und der Bedienungsanleitung verursacht werden, übernimmt die B+B Thermo-Technik GmbH keine Haftung. This device has been designed and tested in accordance to the safety regulations for electronic devices. However, its trouble-free operation and reliability cannot be guaranteed unless the standard safety measures and special safety advises given in this manual will be adhered to when using it.
  • Page 6: Intendend Use

    INSTRUCTION MANUAL Intendend Use The use of the unit in fields other than those indicated under “SAFETY INSTRUCTIONS” is not allowed for safety reasons. This instruction manual does not at all substitute any additional instruction manual of connected accessoriy! Disposal This unit has been marked in accordance with the European Device 2002/96/EC on waste electrical and electronic equipment (WEEE) At the end of its useful operating life, dispose of the unit as electrical scrap.
  • Page 7: Points To Be Observed During Operation

    INSTRUCTION MANUAL 1. Points to be observed during operation a.) Make sure to apply correct operating voltage as low or damaged battery will lead to measuring inaccuracies. If „BAT“ is shown in the display or are the measurements obviously wrong the battery has been used up and needs to be replaced.
  • Page 8: Recalibration

    INSTRUCTION MANUAL 2. Recalibration The measuring device will be calibrated before leaving our works. A recalibration is, therefore, not necessary. If you want to calibrate the device for an existing sensor, please proceed as follows: (calibrate 0°C before scale as otherwise correct adjustment cannot be guaranteed).
  • Page 9: Correctiontable: (Typ. Value)

    INSTRUCTION MANUAL 3. Correctiontable: (typ. value) display tempe- display tempe- display tempe- display tempe- display tempe- display tempe- rature rature rature rature rature rature 1000 1007 1010 1016 1020 1026 1030 1035 1040 1045 1050 1054 1060 1063 1070 1073 1080 1082 1090...
  • Page 10: Specification

    INSTRUCTION MANUAL 4. Specification -50 to +1150°C Range: 1°C Resolution: <1% ±1 digit from -20 to +550°C and 920 to 1150°C, <1,5% ±1 digit from 550 to 920°C. Accuracy: For more detailed values please refer to att. correction table. NiCr-Ni, acc. to ½ DIN 43710 for plug-in operation (not included in scope of supply!) Sensor:: approx.
  • Page 11: Scope Of Delivery

    Articlename Articelnumber Description 0560 1150 1 x Digital Quick-Response Thermometer Digital Quick-Response Thermo- HM 1150 with Operating Instruction on CD-R meter HM 1150 0560 1150-01 1 x Digital Quick-Response Thermometer Digital Quick-Response Thermo- HM 1150 SET meter HM 1150 Set incl.
  • Page 12: General Questions

    INSTRUCTION MANUAL General Questions If you still have questions concerning this product of B+B Thermo-Technik GmbH, please do not hesitate to contact us at: B+B Thermo-Technik GmbH Heinrich-Hertz-Straße 4 D-78166 Donaueschingen Germany Tel.: +49 (0) 771 83160 Fax: +49 (0) 771 831650 E-Mail: info@bb-sensors.com bb-sensors.com We wish you a successful measuring!

Table of Contents