Electrical Installation; Cable Connections; Designation [Models Lt/ G5/ P3/ P7] - B+B Sensors DM-Compact Series Operating Manual

Table of Contents

Advertisement

OPEARTING MANUAL
4

Electrical Installation

4.1

Cable Connections

For the electrical installation of the DM please open at first the cover of the electronic box (4 screws). Below the display are the screw terminals for
the cable connection.
4.1.1

Designation [models LT/ G5/ P3/ P7]

+8..36 VDC
Power supply
GND
Ground (0 V) of power supply
GND
Ground (0 V) of internal in- and outputs
OUT-AMB
Analog output head temperature (mV)
OUT-TC
Analog output thermocouple (J or K)
OUT-mV/mA
Analog output object temperature (mV or mA)
F1-F3
Functional inputs
AL2
Alarm 2 (Open collector output)
3V SW
3 VDC, switchable, for laser-sightingtool
GND
Ground (0 V) for laser-sightingtool
BROWN
Temperature probe head
WHITE
Temperature probe head
GREEN
Detector signal (–)
YELLOW
Detector signal (+)
Technische Änderungen vorbehalten
0141 0315-100
17.08.2015
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
Opened LT/ G5/ P3/ P7 electronic box with terminal connections