Outputs And Inputs; Analog Outputs; Output Channel 1; Output Channel 2 [Lt/ G5/ P3/ P7 Only] - B+B Sensors DM-Compact Series Operating Manual

Table of Contents

Advertisement

OPEARTING MANUAL
5

Outputs and Inputs

5.1

Analog Outputs

The CT has two analog output channels. With the 4M
Model the outputs are freely selectable
5.1.1

Output channel 1

This output is used for the object temperature. The selection of the output signal can be done via the programming keys [►6 Operating]. The
software allows the programming of output channel 1 as an alarm output.
Output signal
Range
Voltage
0 ... 5 V
Voltage
0 ... 10 V
Current
0 ... 20 mA
Current
4 ... 20 mA
Thermocouple
TC J
Thermocouple
TC K
5.1.2

Output channel 2 [LT/ G5/ P3/ P7 only]

The connection pin OUT-AMB is used for output of the head temperature [-20-180 °C or -20-250 °C (on LT21H and LT101H) as 0–5 V or 0–10 V
signal]. The software allows the programming of output channel 2 as an alarm output. Instead of the head temperature T
temperature T
or electronic box temperature T
Obj
Technische Änderungen vorbehalten
0141 0315-100
24.06.2022
ACHTUNG: An die Analogausgänge darf auf keinen Fall eine
Spannung angelegt werden, da dies zur Zerstörung des
Ausgangs führt. Der DM ist kein Zweileitersensor!
Connection pin on CT board
OUT-mV/mA
OUT-mV/mA
OUT-mV/mA
OUT-mV/mA
OUT-TC
OUT-TC
can be selected as alarm source.
Box
B+B Thermo-Technik GmbH | Heinrich-Hertz-Str. 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 831650 | info@bb-sensors.com | bb-sensors.com
According to the chosen output signal different connection
pins on the mainboard are used (OUT-mV/mA or OUT-TC).
also the object
Head
59 / 88