You will find the calibration code on a label fixed on the head cable (near the electronics). Please do not remove this label or make
sure the code is noted anywhere. The code is needed if the electronics must be exchanged or in case of a necessary recalibration of
the sensor.
After you have modified the head code a reset is necessary to activate the change. [►6 Operating]
4.3.2
Sensing Head Cable
On all DM models (exception 3M, P3, P7) the sensing head cable can be shortened if necessary.
On the models 1M, 2M and DM Fast the sensing head cable can be shortened by max. 3 m.
A shortening of the cable will cause an additional measuring error of about 0,1 K/ m.
The 3M models are only available with 3 m cable.