7. Error Messages

If an error occurs during the measurement, the measurement is inter-
rupted and an error message, e.g. «ERR 3», is displayed.
Error
Description Potential cause and remedy
«ERR 1»
Signal too
The pulse signals on the cuff are too weak. Re-
weak
position the cuff and repeat the measurement.*
AN
«ERR 2»
Error signal During the measurement, error signals were
detected by the cuff, caused for instance by
AO
movement or muscle tension. Repeat the
measurement, keeping your arm still.
«ERR 3»
No pressure
An adequate pressure cannot be generated in
in the cuff
the cuff. A leak may have occurred. Check that
AO
the cuff is correctly connected and is not too
loose. Replace the batteries if necessary.
Repeat the measurement.
«ERR 5» Abnormal
The measuring signals are inaccurate and no
result
result can therefore be displayed. Read through
the checklist for performing reliable measure-
ments and then repeat the measurement.*
8
Error
Description Potential cause and remedy
«ERR 6» MAM Mode
There were too many errors during the
measurement in MAM mode, making it
error
impossible to obtain a final result. Read
through the checklist for performing reliable
measurements and then repeat the
measurement.*
«ERR
Wrist Position
Possibly some problem occurs on the sensor
for Wrist Position Indicator. Turn off and
Indicator error
21»
remove the battery then repeat measurement.
If still displayed, please contact your local
Microlife-Service.
«HI»
Pulse or cuff
The pressure in the cuff is too high (over
299 mmHg) OR the pulse is too high (over
pressure too
199 beats per minute). Relax for 5 minutes
high
and repeat the measurement.*
«LO»
Pulse too
The pulse is too low (less than 40 beats per
low
minute). Repeat the measurement.*
* Please immediately consult your doctor, if this or any other
problem occurs repeatedly.
If you think the results are unusual, please read through the
information in «Section 1.» carefully.