Chapter 5 Measurements

5. Measurements

5.1

Measuring an instantaneous value

1
Settings
Example:
Measuring an instantaneous value of three-phase three-wire two-current
(3P3W2I) 200-V line (50 Hz, 120-A load) using the clamp-on probe 96030
(rating: 200 A)
TOP
MENU
ENTER
General 1/2
General 2/2
Save 1/2
Save 2/2
ENTER
2
Wiring
After the above settings have been completed, carry out wiring, referring to
Chapter 4 in this manual.
IM CW240P-E
Displays SETUP.
Using the right and left direction cursor key, display the
General 1/2 screen.
Set up necessary items on the screen.
For details, refer to Chapter 3 in this manual.
Next, using the up and down direction cursor key,
display the General 2/2 screen.
Also set up necessary items on the screen.
For details, refer to Chapter 3 in this manual.
Using the right and left direction cursor key, select Save 1/2 screen.
Next, using the up and down direction cursor key,
select Save 2/2 screen.
This completes the settings, returning you to the Top Menu screen.
Top Menu screen
5-1