Test Equipment Depot CW240 User Manual page 51

Clamp-on power meter
Table of Contents

Advertisement

Symptom
4)
Key operation cannot be done.
5)
Saving and writing to internal
memory cannot be done.
6)
Saving and writing to the
PC card cannot be done.
7)
Communication cannot be done
through the RS-232 interface.
8)Printing cannot be done.
9)
An error occurs in the opening
message.
6-2
Things to Check
• Confirm that key lock is not displayed at the
upper right of the display area.
• Turn the power OFF then ON again. The problem
may be resolved in the opening self-test.
• It is possible that a power supply error, etc.,
occurred while the internal memory was being
accessed.
Format the internal memory in the file processing
mode. Data saved in the internal memory will be
lost.
• Confirm that the PC card has been inserted
correctly.
• Confirm that the PC card has been formatted.
• Confirm that the capacity of the PC card has not
been exceeded.
• Confirm that the communication parameters of
the CW240 and the controller, etc., match.
• Confirm that the specifications of the cables
connecting the CW240 and the controller, etc.,
are suitable for the purpose.
• Confirm that printer power is ON.
(Refer to the printer's instruction manual.)
• Confirm that connection cable specifications
match.
• Confirm that connection cables are connected
correctly.
• Confirm that the communication parameters of
this device and the printer match.
• Confirm that the print media has been set
correctly.
• This is a hardware error. Contact to your local
service representative.
Reference
Section
15.3
3.6
9.4
11.2
9.4
11
6.6
10.2
10.3
10.3
10.3
10.3
3.6
IM CW240P-E

Advertisement

Table of Contents
loading

Table of Contents