70 | Full Leg Full Spine DR Retrofit System | Technical Data

Full Leg Full Spine wall stand technical data

Table 9: Full Leg Full Spine wall stand technical data
Type
Eelectrical connection operating
voltage
Dimensions
(without wall-mounted base)
Weight approx.
(without wall-mounted base)
Front panel of the DR detector hold-
er attenuation
Environmental requirements
Room temperature
Maximum temperature change
Relative humidity
Atmospheric pressure
Relative altitude
Environmental requirements (storage)
Temperature
Environmental requirements (transport)
Temperature
Expected product lifetime
0326A EN 20211014 1141
6005/100
24V DC 2.0A
Width: 735 mm (one handle)
Height: 1374 mm
Depth: 171 mm
25 kg
≤ 1.2 mm Al
According to DIN EN 60601-1-3 with
100kV and HVL 3.6 mm Al, FDA 21
CFR § 1020.30 (n) with 100kV and
HVL 3.6 mm Al
10°C to 35°C
0.5 °C/min.
30% to 75%
70 kPa to 106 kPa
3000 m to 0 m
-20°C to +40°C
-20°C to +40°C
7 years