Execute the exposures

1. Rotate the X-ray tube aligning the X-ray field to the first position of the DR
detector as indicated on the front panel of the DR detector holder.
2. Press the exposure button to acquire the first partial image.
The partial image is sent to the NX workstation.
3. Press the button on the Full Leg Full Spine control unit.
The DR detector is moved to the next position inside the DR detector
holder.
4. Repeat previous steps to move the DR detector to each required position,
rotate the X-ray tube correspondingly and acquire the partial images.
If only two partial images are required and the examination contained
three partial image thumbnails, the FLFS workflow can be cancelled after
the second exposure. The stitching must be started manually.