AGFA 6005/100 6005/200 User Manual page 74

Full leg full spine dr retrofit system
Table of Contents

Advertisement

74 | Full Leg Full Spine DR Retrofit System | Technical Data
with IEC
610004-2
Fast transient
electrical dis-
turbance varia-
bles / bursts in
accordance with
IEC 610004-4
Impulse voltag-
es (surges) in
accordance with
IEC 61000-4-5
Voltage break-
throughs, short
term interrup-
tions and varia-
tions in the volt-
age supplied in
accordance with
IEC 61000-4-11
Magnetic field
at the supply
frequency
(50/60 Hz) in
0326A EN 20211014 1141
+ 15 kV air dis-
charge
+ 1 kV mains
+ 0,5 kV data lines
+ 1 kV push-pull
voltage
+ 2 kV common
mode voltage
• 0% U
for ½ peri-
r
od
• 0% U
for 1 peri-
r
od
• 70% U
(30%
r
breakthrough of
U
) for 25 peri-
r
ods
• 0% U
for 250 pe-
r
riods
30 A/m
+ 15 kV air dis-
ramic tiles. The
charge
relative humidi-
ty must be at
least 30%, if the
floor is made of
synthetic materi-
al.
+ 2 kV mains
The quality of
the voltage sup-
+ 1 kV data
plied should cor-
lines
respond to a typ-
ical commercial
or clinical envi-
ronment.
+ 1 kV push-
The quality of
pull voltage
the voltage sup-
plied should cor-
+ 2 kV common
respond to that
mode voltage
of a typical com-
mercial or clini-
cal environment.
The quality of
• 0% U
for ½
r
the voltage sup-
period
ply should corre-
• 0% U
for 1
r
spond to that of
periods
a typical com-
• 70% U
r
mercial or clini-
(30% break-
cal environment.
through of
U
) for 25
If the user wants
r
periods
the device to
work continu-
• 0% U
for
r
ously, even
250 periods
when the energy
supply is inter-
rupted, it is rec-
ommended to
use an energy
supply free of in-
terruptions or a
battery.
30 A/m
Magnetic field at
the network fre-
quency should
correspond to
the typical val-