Emerson Micro Motion 5700 Instruction Manual page 29

Transmitters with configurable inputs and outputs
Table of Contents

Advertisement

Installation Manual
MMI-20027478
4.4.1
Wire the mA/HART
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-6: mA/HART output wiring (internally powered)
A
A. mA/HART output
B. 250–600 Ω resistance
C. HART device
4.4.2
Wire the mA/HART
Procedure
Wire to the appropriate output terminal and pins.
Figure 4-7: mA/HART output wiring (externally powered)
A
A. mA/HART output
B. 5–30 VDC (maximum)
C. 250–600 Ω resistance (see
D. HART device
Installation Manual
1
2
1
2
®
output (internally powered)
®
output (externally powered)
Figure 4-8
for maximum loop resistance)
Wiring the channels
April 2022
C
B
B
C
D
29