Operating conditions: 10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity
Storage conditions:
-20 - +55 °C / -4 - +131 °F
15 - 95 % relative maximum humidity
Weight:
130 g (including batteries)
Dimensions:
80 x 70 x 70 mm
Cuff size:
13.5 - 21.5 cm (5.25 - 8.5 inches)
Measuring procedure: oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic
Measurement range:
20 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display
range:
0 - 299 mmHg
Resolution:
1 mmHg
Static accuracy:
pressure within ± 3 mmHg
Pulse accuracy:
± 5 % of the readout value
Voltage source:
2 x 1.5 V alkaline batteries; size AAA
Battery lifetime:
approx. 460 measurements
(using new batteries)
BP W100
IP Class:
IP20
Included accessories: blood pressure monitor, storage case, 2 x
AAA batteries, instruction manual, guar-
antee card, blood pressure diary
Reference to
EN 1060-1 /-3 /-4; IEC 60601-1;
standards:
IEC 60601-1-2 (EMC); IEC 60601-1-11
Expected service life: Device: 5 years or 10000 measurements
Accessories: 2 years
This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.
Date of production: first four digits of the serial number of the device.
First and second digit – the week of the year / third and fourth digit –
the year of production.